Tests For: Total aluminium in natural and treated waters
Test Range: 0–0.5 mg/L
Reagent Chemistry Used: Ericochrome cyanine R
Basis of Test Method: Standard Method 3500-B, ISO15923-2:2017
Method Detection Limit*: 0.01 mg/L
Limit of Quantification**: 0.04 mg/L
Aluminium sulphate is widely used as a coagulant in drinking water treatment. The determination of aluminium (residual alum) is usually required for the control of alum coagulation and filtration processes at water works.
Aluminium salts are found in natural waters; levels are reported to be increasing particularly in areas affected by acid rain. High aluminium levels can be toxic to fish and aquatic life. Aluminium determination is necessary therefore for environmental control and for testing water used for fish farms, etc.
The Palintest Aluminium test provides a simple method of measuring aluminium levels in natural and drinking waters over the range 0 – 0.5 mg/l.
Aluminium reacts with Eriochrome Cyanine R indicator in slightly acid solution to produce a pink-red coloured complex. In the Palintest Aluminium method the necessary reagents are incorporated into two test tablets. The test is simply carried out by adding one of each tablet to a sample of the water. The first tablet acidifies the sample to bring any colloidal aluminium into solution and the second tablet buffers the solution to provide the correct conditions for the test.
The intensity of the colour produced in the test is proportional to the aluminium concentration and is measured using a Palintest Photometer.
The presence of ascorbic acid eliminates interference from iron and manganese.
The presence of polyphosphate or fluoride can lead to low aluminium readings. Polyphosphate is unlikely to be present in significant quantities in normal water samples. Fluoride will only be significant for control samples from water works where fluoridation is practised. In such cases samples should preferably be taken before the final fluoridation stage.
For samples taken after fluoridation such as those from water distribution systems, or for samples containing natural fluoride, the aluminium concentration should be corrected. To obtain the corrected aluminium concentration use the following formula:
Corrected Value = Photometer Reading x (1 + (0.4 x Fluoride Concentration))
A separate test is available for measuring fluoride.