Products
Kemio™
Photometers
Wagtech™
Siren
BW
Visual
Electrochemical Meters
Applications
Water Utilities
Pool & Spa
Humanitarian
Food & Beverage
View All Applications
Parameters
After Sales
Accessories and Check Standards
FAQs
Product Downloads
Product Registration via Palintest Connect
Reagents and Sensors
Safety Data Sheets
Servicing and Calibration
Technical Support
Content Hub
Content Hub
Webinars
About Us
Purpose, Vision, Mission & Values
Purpose, Vision, Mission & Values
Contact Us
Search
Search
Search
Looking for Reagents?
Looking for Accessories?
Looking for Technical Support?
MENU
EN
Products
Kemio™
Photometers
Wagtech™
Siren
BW
Visual
Electrochemical Meters
Applications
Water Utilities
Pool & Spa
Humanitarian
Food & Beverage
View All Applications
Parameters
After Sales
Accessories and Check Standards
FAQs
Product Downloads
Product Registration via Palintest Connect
Reagents and Sensors
Safety Data Sheets
Servicing and Calibration
Technical Support
Content Hub
Content Hub
Webinars
About Us
Purpose, Vision, Mission & Values
Purpose, Vision, Mission & Values
Contact Us
Search
Home
>
Tests
>
Phosphorous, Total/12
Phosphorous, Total/12
Photometric test for Total Phosphorous in natural water and wastewater
Colour Change
Colourless to Blue
Range:
0 - 12 mg/L P
Technical Resource:
Phosphorous, Total/12 Test Method, Technical Information
01
Turn on heater block and set temperature to 105⁰C. Note: Timer may need setting above zero for warm up to start.
02
Add 2 mL of sample to a Tubetest.
03
Add two Digest Ox tablets and crush and stir.
04
Set timer on heater block to 60 minutes.
05
Place the tube in the heating block.
06
Ensure safety screen is in position and allow digestion to take place for 60 minutes.
07
Carefully remove tube and place in a rack to cool.
08
Fit the Tubetest adapter to Lumiso. Then, place an unused Total Phosphorous Tubetest in cell holder and press Blank.
09
Removed the digested tube from the rack and add 2 mL of PhosNeut Solution.
10
Continue with the same tube by adding a Tubetest Phos No1 tablet. Crush and stir, until fully dissolved.
11
Then, add a Tubetest Phos No2 tablet and crush and stir.
12
Ensure cap is tight, then invert tube a few times to mix.
13
Place in the cell holder and press Measure.
14
Allow timer to count down 10 minutes. The concentration of Total Phosphorous will then be displayed